6003A Firmware Update

6003A Firmware Update (.zip)

TITLE USE WITH
6003A All

The updating of embedded software should only be performed after given specific instructions to do so from Technical Support or a Fluke Calibration Authorized Service Center.

If the update isn't performed properly, your product may be inoperable until it is.

Instructions to update by USB

Installation

Put the CD into the computer, wait for autorun and select “Software installation” and “Uploader” or start direct “install\freeware\uploader”.

Uninstall Uploader

To remove the Uploader software from your system, you should run “uninstall” from “Programs” ("Start -> Programs -> Uploader -> Uninstall Uploader" from the Windows Start menu).

Uploading Firmware

1. Connect the calibrator to the computer using an RS232 cable.

2. Select RS232 in the instrument menu as the active interface and set the baud rate to 9600 Bd.

3. Start the Uploader “Start->Program files->Uploader->Uploader.exe”. The following screen will appear:

Step 3 Uploader

4. Select computers port (COM1 … COM32) and the source file. The source file is a data file that contains the new firmware with the extension “.upl”.

5. Press “Connect”. The following screen will appear:

Step 5 Connection screen

6. It is recommended to keep the transfer speed as-is and press “Upload” (in the case of an unsuccessful upload, reduce the transfer speed and try again). The following screen will appear:

Step 6 Transfer Progress

7. Uploader automatically reloads all internal programs step by step.

8. After successful upload, the Uploader displays the following:

Step 8 Success Screen

Problems While Uploading New Firmware

In the case that the upload does not finish properly (power supply disconnection, failed communication, etc.) the instrument will no longer contain internal software. Switch the instrument’s power supply off, wait approximately 10 seconds, and switch it on again. The instrument will display the message “Internal SW damaged!”. Repeat the procedure upload. The instrument is switched to RS232, 9600Bd mode automatically.

Updates Include:

Version 1.048 (2019 Oct. 25)

Fix: Bug fixed that caused the GUI to lock up when too many characters were entered into the calibration password field.