5606 Full Immersion PRT

  • 5606 Full Immersion PRT 1

主要特性

  • 可在 -200 °C 至 160 °C 的温度范围内运行。
  • 准确度 ±0.05 °C。
  • 这些 PRT 的过渡连接处和引线均能够承受全量程的 PRT 温度。
  • 独特的探头密封可防止湿气进入。
  • 提供 NVLAP 认可的校准。
  • 适用于校准/验证实验室或生物冷冻箱、高压灭菌器、烤箱、试验箱、温度受控区等设备中进行的 IQ/OQ/PQ 认证程序、温度测绘或数据记录中的传感器。

产品概述: 5606 Full Immersion PRT

The 5606 is just two inches (50 mm) in length with a sheath diameter of 1/8 inch (3.1 mm). Since it can be fully immersed over its entire temperature range, you don’t have to worry about calculating minimum immersion depth— immerse the entire probe, transition junction, and lead wires, in either non-corrosive liquids or dry mediums ranging from –200 °C to 160 °C.

The 100 Ω precision sensing element (alpha = 0.00385) is specially constructed to prevent the ingress of moisture while achieving calibration accuracies as good as ± 0.05 °C. You get both accuracy and flexibility built into a small, durable package.

The lead wires of the 5606 are made of single-conductor enameled copper wire which prevents moisture from wicking to the sensing element as it would if a more common, stranded lead wire were used. The diameter of the lead wire is no greater than 0.2 mm, making it easy to run lead wires through a freezer door to your readout or transmitter with effectively zero energy loss. The 5606 has been proven in common heat transfer fluids such as silicone oil, mineral oil, ethanol, and even liquid nitrogen with no degradation of the transition junction seal or lead wire insulation.

The 5606 Full Immersion PRTs are designed to perform in extreme environments where both the transition junction and the lead wires are required to withstand temperatures covering the entire operating range of the probe. (“Immersion" in this case simply means exposure to temperatures other than ambient and does not necessarily refer to the medium to which the PRT and its components are subjected.)

Applications where “full immersion" of probe, transition junction, and lead wire is required can include calibration or validation of sensors used in laboratory or bio freezers, walk-in refrigerators, autoclaves, ovens, stability test chambers, furnaces or incubators. Other applications might include IQ/OQ/PQ qualification procedures, temperature mapping, or data logging in temperature controlled spaces or environmental chambers.

产品规格: 5606 Full Immersion PRT

General Specifications
Temperature range-200°C to 160°C
Nominal resistance at 0.01°C100 Ω ±0.1 Ω
Temperature coefficient0.00385 Ω/Ω/°C
Accuracy1±0.04°C at 0.01°C
±0.06°C at 160°C
Short-term repeatability2±0.03°C at 0.01°C
±0.04°C at 160°C
Drift3±0.03°C at 0.01°C
±0.04°C at 160°C
Hysteresis±0.015°C
Sheath length50 mm ±5 mm
(2 in ±0.2 in)
Sheath diameter3.1 mm ±0.1 mm (1/8 in ±0.004 in)
Sheath material316 SST
Transition junction temperature range4-200°C to 160°C
Transition junction dimensionsNo transition
Sensor length30 mm ±3mm (1.2 in ±0.1 in)
Sensor location3 mm ±1 mm from tip (1.2 in ±0.04 in)
Minimum insulation resistance20 MΩ at 23°C
Minimum immersion lengthFull immersion
Maximum immersion depth in liquid mediumFull immersion
Maximum immersion depth in dry MediumFull immersion
Response time512 seconds typical
Self heating (in 0°C bath)±0.003°C
Lead-wire cable typeEnameled copper wire
Lead-wire length2.4 meters (8 ft)
Lead-wire temperature range160°C
CalibrationCalibration not included. NVLAP accredited calibration available. See ordering information.
1. Includes calibration and 100 hr drift (k = 2).
2. Three thermal cycles from min to max temp, includes hysteresis, 95% confidence (k = 2).
3. After 100 hours at max temp, 95% confidence (k = 2).
4. Temperatures outside this range will cause irreparable damage.
5. Per ASTM E 644.
Calibration Uncertainty
Temperature-197°C1924 Calibration Uncertainty6 0.03°C
Tolerance w/out Calibration7 0.4°C
 -38°C1924 Calibration Uncertainty6 0.03°C
Tolerance w/out Calibration7 0.2°C
 0°C1924 Calibration Uncertainty6 0.03°C
Tolerance w/out Calibration7 0.1°C
 100°C1924 Calibration Uncertainty6 0.045°C
Tolerance w/out Calibration7 0.3°C
 157°C1924 Calibration Uncertainty6 0.05°C
Tolerance w/out Calibration7 0.4°C
6. Lab code 200348-0
7. Each PRT includes a certificate with a measured R0 value.
An uncalibrated 5606 meets ASTM E 1137 Grade A Classification of Tolerances.
Use the coefficients below to achieve the tolerances listed in the “No Calibration” column or calculate the Grade A tolerance using the following formula:
Grade A = ±[0.13 + 0.0017|t|]°C
where: |t| = value of temperature (°C) without regard to sign. See ASTM E 1137.
R0: Enter the R0 value provided
A: 3.9083 x 10–3
B: –5.775 x 10–7
C: –4.183 x 10–12

型号: 5606 Full Immersion PRT

5606-50-P

Fluke 5606-50-P Full Immersion PRT
Full Immersion PRTs Probe, 50 mm, –200 to 160°C, INFO-CON termination for 1523 or 1524

For Fluke 5606-50-X with other terminations below, please visit us.Flukecal.com and request a quote


† X = termination. Specify “A” (INFO-CON for 914X), “B” (bare wire), “D” (5-pin DIN for Tweener Thermometers), “G” (gold pins), “I” (INFO-CON for 1521 or 1522), “J” (banana plugs), “L” (mini spade lugs), “M” (mini banana plugs), or “S” (spade lugs).

5606 Full Immersion PRT

1930-4-10
系统对比校准,-200 至 157°

5606-50-X
探头,浸入式 PRT,50 mm,-200 到 160°C

† X = 终止。指定“A”(适用于 914X 的 INFO-CON)、“B”(裸线)、“D”(适用于 Tweener 温度计的 5 针 DIN)、“G”(金针)、“J”(香蕉插头)、“L”(微型平接线片)、“M”(微型香蕉插头)、“P”(适用于 1523 或 1524 的 INFO-CON)或“S” (平接线片)。
1930-5-10

系统对比校准,-40 至 157