5960A Triple Point of Argon System

  • 5960A Triple Point of Argon System

主要特性

  • 0.25 mK 的低不确定度和超过 30 小时的长温坪。
  • 四个进入井,径向均匀度小于 0.05 mK
  • 总浸深 480 mm。
  • 纯度 99.9999% 的氩气。
  • 支持多语言显示。
  • 可复现氩三相点的一致、可重现过程。

产品概述: 5960A Triple Point of Argon System

Low calibration uncertainty


A number of technology advancements are designed into the Fluke Calibration 5960A system to achieve a low uncertainty of 0.25 mK. For example, the argon cell is carefully built with an ultra-clean process using 99.9999 % (six nines) purity argon resulting in an uncertainty factor from impurities of 0.015 mK. The deep, 480 mm immersion depth of the entrant well minimizes the effects of stem conduction to less than 0.01 mK and radial uniformity to less than 0.05 mK.

High lab throughput


The 5960A system utilizes liquid nitrogen (LN2) in a Dewar vessel surrounding an argon cell to achieve –189.3442 °C. A digital controller combined with a heater maintains a temperature plateau lasting up to 30 hours with 0.1 mK variation without using complicated pressure controls. By regular renewal of liquid nitrogen in the system, the plateau length can even be extended to four days. The system’s long, stable temperature plateau lets you carry out your SPRT calibrations with better quality and more efficiency. Four entrant wells increase lab productivity by allowing you to calibrate four SPRTs simultaneously.

Ease of operation and maintenance


The display panel of the 5690A provides a simple menu structure with soft function keys for setup and programming the argon system. The readout displays simultaneously the set-point temperature, actual control sensor temperature and percent heater power. The display is selectable in eight languages—English, French, Spanish, Italian, German, Russian, Chinese, and Japanese. Applying positive pressure with helium or dry nitrogen to an entrant well port prevents your SPRTs from being bound in the entrant well by freezing moisture. You can insert and remove SPRTs as desired from the system over the duration of a single plateau. Other systems require that you first melt the cell in order to add or remove an SPRT.

产品规格: 5960A Triple Point of Argon System

Specifications
Assigned value-189.3442 °C
Argon gas purity99.9999 % (6N purity)
Uncertainty (k=2)0.25 mK
Length of plateau (0.1 mK)>30 hours
Argon cell depth*160 mm
Number of entrant wells4
Radial uniformity0.05 mK
Total immersion depth480 mm
Entrant well inner diameter (ID)8.0 mm
Argon gas volume13.4 moles (535 grams)
Liquid nitrogen Dewar volume44.2 liters
Display resolution0.001 °C
Set-point accuracy0.1 °C (adjustable)
Power requirements100 V to 115 V (± 10 %) 50/60 Hz, 230 W 230 V (± 10 %) 50/60 Hz, 230 W
System fuse ratings 115 V: 2 A T 250 V
230 V: 1 A T 250 V 
Size (HxWxD) 952 mm x 673 mm x 483 mm (37.5 in x 26.5 in x 19 in) 
Weight 94 kg (207 lbs) 
*Measured from the bottom of the probe well to the surface of argon sample

型号: 5960A Triple Point of Argon System

5960A
Triple Point of Argon System